Gungor, GorkemKorkmaz, LatifKayalilar, NamikAydin, GokhanYapici, BulentMustafayev, Teuta ZotoAtalar, BanuOzyar, Enis2023-02-212023-02-212020-01-0110.7759/cureus.7334https://hdl.handle.net/11443/1932http://dx.doi.org/10.7759/cureus.7334Purpose To evaluate the intensity modulated radiotherapy (IMRT) quality assurance (OA) results of the multichannel film dosimetry analysis with single scan method by using Gafchromic (TM) EBT3 (Ashland Inc., Covington, KY, USA) film under 0.35 T magnetic field. Methods Between September 2018 and June 2019, 70 patients were treated with ViewRay MRldian (R) (ViewRay Inc., Mountain View, CA) linear accelerator (Linac). Film dosimetry OA plans were generated for all IMRT treatments. Multichannel film dosimetry for red, green and blue (RGB) channels were compared with treatment planning system (TPS) dose maps by gamma evaluation analysis. Results The mean gamma passing rates of RGB channels are 97.3\% +/- 2.26\%, 96.0\% +/- 3.27\% and 96.2\% +/- 3.14\% for gamma evaluation with 2\% DD/2 mm distance to agreement (DTA), respectively. Moreover, the mean gamma passing rates of RGB channels are 99.7\% +/- 0.41\%, 99.6\% +/- 0.59\% and 99.5\% +/- 0.67\% for gamma evaluation with 3\% DD/3 mm DTA, respectively. Conclusion The patient specific QA using Gafchromic (TM) EBT3 film with multichannel film dosimetry seems to he a suitable tool to implement for MR-guided IMRT treatments under 0.35 T magnetic field. Multichannel film dosimetry with Gafchromic (TM) EBT3 is a consistent QA tool for gamma evaluation of the treatment plans even with 2\% DD/2 mm DTA under 0.35 T magnetic field presence.multichannel film dosimetryimrtpatient qamrigrtviewray0.35 tebt3Multichannel Film Dosimetry for Quality Assurance of Intensity Modulated Radiotherapy Treatment Plans Under 0.35 T Magnetic FieldArticleWOS:000521215700001