<link rel="stylesheet" href="styles.f3b1fba60ec7970c.css">

Closed-loop synchronization versus conventional synchronization in spontaneously breathing pediatric patients (CHESTSIPP) - a randomized controlled cross-over study

dc.contributor.authorCeylan, Gokhan
dc.contributor.authorSarac, Ozlem
dc.contributor.authorAtakul, Gulhan
dc.contributor.authorSoydan, Ekin
dc.contributor.authorColak, Mustafa
dc.contributor.authorTopal, Sevgi
dc.contributor.authorHepduman, Pinar
dc.contributor.authorNovotni, Dominik
dc.contributor.authorMeyer, Johannes
dc.contributor.authorKaraaslan, Utku
dc.contributor.authorRevely, Jean-Pierre
dc.contributor.authorAgin, Hasan
dc.date.accessioned2026-11-05T11:06:33Z
dc.date.issued2026
dc.description.abstractBackground: Patient-ventilator asynchrony (PVA) is common during invasive mechanical ventilation in children and may increase respiratory workload and discomfort. The IntelliSync+ (IS+) algorithm provides closed-loop synchronization by continuously analyzing airway pressure and flow waveforms to optimize inspiratory and expiratory cycling in real time. This study evaluated the efficacy and safety of IS+ compared with conventional physician-tailored synchronization in spontaneously breathing pediatric patients. Methods: Multicenter, prospective, randomized controlled, single-blind crossover trial conducted in four pediatric intensive care units between March 2024 and May 2025. Patients aged 1 month to 18 years undergoing invasive mechanical ventilation were randomized to begin with either IS+ or conventional synchronization. Each mode was applied for 90 min (30-min run-in and 60-min measurement), separated by a 30-min washout period. The primary outcome was the Asynchrony Index (AI). Secondary outcomes included major and minor asynchrony indices, Comfort-B scores, end-tidal carbon dioxide (EtCO2), oxygen saturation (SpO(2)), and leak percentage. Findings: Twenty-five patients completed both study phases. Compared with conventional synchronization, IS+ significantly reduced the overall AI (median 5.1% [IQR 4.2-6.4] vs. 12.4% [IQR 9.9-15.6]; P < 0.001), as well as both major and minor asynchrony indices (P < 0.001 for both). The frequencies of double triggering, auto-triggering, trigger delay, and early cycling decreased significantly, while ineffective efforts were unchanged and late cycling slightly increased. Comfort-B scores improved modestly (P = 0.02) and EtCO2 decreased (P < 0.001), whereas SpO(2) and leak remained stable. Conclusion: Closed-loop synchronization with IS+ significantly improved patient-ventilator interaction without compromising comfort, oxygenation, or safety. These results support closed-loop synchronization as a feasible and physiologically sound strategy for improving ventilatory support quality in pediatric invasive mechanical ventilation.
dc.description.pages1745939
dc.description.sponsorshipThe author(s) declared that financial support was not received for this work and/or its publication.
dc.description.volume13
dc.identifier.doi10.3389/fmed.2026.1745939
dc.identifier.issn2296-858X
dc.identifier.urihttps://doi.org/10.3389/fmed.2026.1745939
dc.identifier.urihttps://openaccess.acibadem.edu.tr/handle/11443/11363
dc.identifier.wosWOS:001735127200001
dc.language.isoen
dc.publisherFRONTIERS MEDIA SA
dc.relation.hasversion10.3389/fmed.2026.1745939
dc.relation.ispartofFRONTIERS IN MEDICINE
dc.subjectasynchronies
dc.subjectclosed-loop
dc.subjectmechanical ventilation
dc.subjectpediatric acute respiratory distress syndrome (PARDS)
dc.subjectspontaneous breathing
dc.subjectsynchronization
dc.subjectwaveform analysis
dc.subjectVENTILATOR ASYNCHRONY
dc.subjectPRESSURE-SUPPORT
dc.subjectVOLUME
dc.subjectMedicine, General & Internal
dc.subjectGeneral & Internal Medicine
dc.titleClosed-loop synchronization versus conventional synchronization in spontaneously breathing pediatric patients (CHESTSIPP) - a randomized controlled cross-over study
dc.typeArticle
oaire.citation.editionScience Citation Index Expanded (SCI-EXPANDED)
oaire.citation.volume13
person.identifier.ridAFW-8030-2022
person.identifier.ridJAO-2810-2023
person.identifier.ridJYQ-2668-2024
person.identifier.ridJLL-9304-2023
person.identifier.ridJDW-3813-2023
person.identifier.ridAAG-3609-2021
person.identifier.ridNHP-1194-2025

Files