<link rel="stylesheet" href="styles.f3b1fba60ec7970c.css">

Wearable Assessment of Dynamic Trunk Sway Reveals Directional Balance Adaptations During Sling-Assisted Walking After Stroke

dc.contributor.authorYalcin, Begum
dc.contributor.authorGokhan, Yigit Can
dc.contributor.authorSirzai, Hulya
dc.contributor.authorYavuzer, Gunes
dc.contributor.authorArgunsah, Hande
dc.date.accessioned2026-11-05T11:00:29Z
dc.date.issued2026
dc.description.abstractBackground: Quantitative assessment of dynamic balance remains challenging in clinical practice. Wearable inertial measurement unit (IMU)-based technologies offer an objective and accessible approach for monitoring postural control. This study aimed to develop and preliminarily validate a wearable IMU-based trunk sway monitoring system (SwayTracker) and investigate trunk sway characteristics in stroke patients walking with and without arm sling support. Methods: A sternum-mounted IMU system was developed to quantify dynamic trunk sway during walking. Fifteen healthy adults and fourteen stroke patients participated. The healthy participants established normative reference values, whereas the stroke patients completed walking trials with and without arm sling support. Trunk sway was quantified using anteroposterior (AP) and mediolateral (ML) deviations and a polar-coordinate-based sway model. Results: The healthy reference cohort exhibited a mean trunk sway magnitude (radius) of 8.20 +/- 3.16 degrees. The stroke patients demonstrated greater sway during unsupported (15.36 +/- 5.83 degrees) and sling-assisted walking (15.35 +/- 4.59 degrees). Although overall sway magnitude remained unchanged, the mean sway direction shifted from 69.49 degrees to 110.20 degrees, indicating a redistribution of trunk sway from the anterior-right toward the anterior-left quadrant. Forward sway remained the dominant AP component, whereas ML sway shifted from predominantly rightward to leftward with sling use. Conclusions: SwayTracker provides a feasible method for objective assessment of dynamic trunk sway during walking. The stroke patients exhibited increased sway magnitude and altered directional organization compared with healthy individuals. Arm sling use primarily modified ML postural compensation patterns rather than reducing overall trunk sway, highlighting the potential of wearable trunk sway monitoring for gait and balance assessment in neurological rehabilitation.
dc.description.issue15
dc.description.pages6104
dc.description.sponsorshipThis research was funded by TUBITAK (grant number 125E855).
dc.description.volume15
dc.identifier.doi10.3390/jcm15156104
dc.identifier.issn2077-0383
dc.identifier.urihttps://doi.org/10.3390/jcm15156104
dc.identifier.urihttps://openaccess.acibadem.edu.tr/handle/11443/10454
dc.identifier.wosWOS:001846978000001
dc.language.isoen
dc.publisherMDPI
dc.relation.fundingTUBITAK(Turkiye Bilimsel ve Teknolojik Arastirma Kurumu (TUBITAK))
dc.relation.fundingTUBITAK [125E855]
dc.relation.hasversion10.3390/jcm15156104
dc.relation.ispartofJOURNAL OF CLINICAL MEDICINE
dc.subjecttrunk sway
dc.subjectdynamic balance
dc.subjectstroke rehabilitation
dc.subjectarm sling
dc.subjectpostural control
dc.subjectneurological rehabilitation
dc.subjectGAIT
dc.subjectMedicine, General & Internal
dc.subjectGeneral & Internal Medicine
dc.titleWearable Assessment of Dynamic Trunk Sway Reveals Directional Balance Adaptations During Sling-Assisted Walking After Stroke
dc.typeArticle
oaire.citation.editionScience Citation Index Expanded (SCI-EXPANDED)
oaire.citation.issue15
oaire.citation.volume15
person.identifier.ridNWI-0772-2025
person.identifier.ridHJP-0324-2023

Files

Collections