The EuBIVAS: Within- and Between-Subject Biological Variation Data for Electrolytes, Lipids, Urea, Uric Acid, Total Protein, Total Bilirubin, Direct Bilirubin, and Glucose

dc.contributor.authorAarsand, Aasne K.
dc.contributor.authorDiaz-Garzon, Jorge
dc.contributor.authorFernandez-Calle, Pilar
dc.contributor.authorGuerra, Elena
dc.contributor.authorLocatelli, Massimo
dc.contributor.authorBartlett, William A.
dc.contributor.authorSandberg, Sverre
dc.contributor.authorRoraas, Thomas
dc.contributor.authorCeriotti, Ferruccio
dc.contributor.authorSolvik, Una Orvim
dc.contributor.authorSylte, Marit Sverresdotter
dc.contributor.authorCoskun, Abdurrahman
dc.contributor.authorSerteser, Mustafa
dc.contributor.authorUnsal, Ibrahim
dc.contributor.authorTosato, Francesca
dc.contributor.authorPlebani, Mario
dc.contributor.authorJonker, Niels
dc.contributor.authorBarla, Gerhard
dc.contributor.authorCarobene, Anna
dc.contributor.authorChem, European Federation Clinical
dc.date.accessioned2023-02-21T12:41:18Z
dc.date.available2023-02-21T12:41:18Z
dc.date.issued2018-01-01
dc.description.abstractBACKGROUND: The European Federation of Clinical Chemistry and Laboratory Medicine European Biological Variation Study (EuBIVAS) has been established to deliver rigorously determined data describing biological variation (BV) of clinically important measurands. Here, EuBIVAS-based BV estimates of serum electrolytes, lipids, urea, uric acid, total protein, total bilirubin, direct bilirubin, and glucose, as well as their associated analytical performance specifications (APSs), are presented. METHOD: Samples were drawn from 91 healthy individuals (38 male, 53 female
dc.description.abstractage range, 21-69 years) for 10 consecutive weeks at 6 European laboratories. Samples were stored at -80 degrees C before duplicate analysis of all samples on an ADVIA 2400 (Siemens Healthineers). Outlier and homogeneity analyses were performed, followed by CV-ANOVA on trend-corrected data, when relevant, to determine BV estimates with CIs. RESULTS: The within-subject BV (CVI) estimates of all measurands, except for urea and LDL cholesterol, were lower than estimates available in an online BV database, with differences being most pronounced for HDL cholesterol, glucose, and direct bilirubin. Significant differences in CVI for men and women/women <50 years of age were evident for uric acid, triglycerides, and urea. The CVA obtained for sodium and magnesium exceeded the EuBIVAS-based APS for imprecision. CONCLUSIONS: The EuBIVAS, which is fully compliant with the recently published Biological Variation Data Critical Appraisal Checklist, has produced well-characterized, high-quality BV estimates utilizing a stringent experimental protocol. These new reference data deliver revised and more exacting APS and reference change values for commonly used clinically important measurands, thus having direct relevance to diagnostics manufacturers, service providers, clinical users, and ultimately patients. (c) 2018 American Association for Clinical Chemistry.
dc.description.issue9
dc.description.issueSEP
dc.description.pages1380-1393
dc.description.volume64
dc.identifier.doi10.1373/clinchem.2018.288415
dc.identifier.urihttps://hdl.handle.net/11443/2706
dc.identifier.urihttp://dx.doi.org/10.1373/clinchem.2018.288415
dc.identifier.wosWOS:000448300800017
dc.publisherAMER ASSOC CLINICAL CHEMISTRY
dc.relation.ispartofCLINICAL CHEMISTRY
dc.titleThe EuBIVAS: Within- and Between-Subject Biological Variation Data for Electrolytes, Lipids, Urea, Uric Acid, Total Protein, Total Bilirubin, Direct Bilirubin, and Glucose